Operating Manual V 1.1 – 05/2007
6.
Installation
6.1
Site Requirements
The place of installation must meet the following requirements:
No direct sunlight.
Power supply at a distance no greater than approx. 3 m.
No drafts (air condition outlets etc.).
Even floor.
Vibration-free floor.
The instrument is designed for indoor use only.
Obstruction-free access to the handwheel.
The power switch/circuit breaker must be freely and easily accessible.
Room temperature always approx. 22 °C.
The relative humidity must not exceed 60 %.
To ensure proper function of the instrument, it must be set up while
maintaining a minimum distance from walls and furniture.
Distance to walls and furniture, calculated from the cabinet:
- rear:
15 cm
- right side:
30 cm
- left side:
15 cm
No heat dissipating appliances around.
High room temperatures and excessive air humidity affect the
cooling capacity of the cryostat and lead to ice forming in the
instrument!
Plug the instrument into power sockets with ground only.
Power cable length:
up to 3.5 m
Extendable:
no