Leica CM1950 Manuel D’Utilisation

Page de 74
21
Leica CM1950  – Cryostat
6.
Installation
6.2
Transport to the Site
First, check if the location meets the conditions specified in "Site
requirements".
Transport the instrument to the desired location.
Observe the following:
The instrument must be transported in an upright position or slight-
ly tilted (max. 30°)!
When transporting the instrument on wheels
(2) grip the cabinet only at the marked loca-
tions ( ).
To do so, unscrew the adjustable feet using
the No. 13 open-end wrench (when subse-
quently transporting the instrument on cas-
tors, screw the feet back in as far as they will
go). To ensure a secure upright position at its
intended location, align both adjusting feet (1).
1
When tilting the instrument 2 people must counterbalance from the
front side to prevent the instrument from falling down and causing
severe injury!
1
2
2
2 (covered)
Fig. 12
Before transport or relocation, remove
the filter bag from the chamber. If you
fail to do this, the filter bag will thaw,
then freeze solid when the instrument
is reconnected. When subsequently
removed, the filter will be destroyed,
causing section waste to get into the
bacteria filter. (Refer to p. 26.)
When not using the extraction for a long
period, tightly close the opening for the
extraction hose using the silicone stop-
per (5, page 28) included in the scope of
delivery.