4. CONNECTIONS  4.2
 Cautions on Connections
 
 
Observe the following cautions during connections for securing safety and reliability. 
 
  1
    Power supply
 
Use a single-phase power supply having a stable 
voltage without any waveform distortion for the 
purpose of preventing wrong operations. 
 
 
Prepare a switch and an overcurrent 
protective device (3 A) to the power 
supply for preventing an accidental 
electric shock during connection work. 
This instrument is not provided with any 
replaceable overcurrent protective device. 
 
 
 
Make sure to turn off the power supply 
before connecting the power and the 
input/output terminals to prevent an 
electric shock. 
 
 
  2
  Keep the input/output connections 
away from high voltage power 
circuits
 
Don’t place the input/output cables close to or in 
parallel with any strong power circuits including 
power lines. Place the cables 50cm or more away 
from high voltage power circuits when they are 
placed close to or in parallel to other circuits. 
 
  3
  Keep the thermocouple input away 
from a heat source 
For thermocouple inputs, keep the input terminals 
away from a heat source (a heating body) to  
reduce a reference junction compensation error. 
Don’t expose the input terminals to direct sunlight, 
etc. 
 
4
  Keep the input/output connections 
away from noise source
 
Keep all connection cables away from noise 
source as far as possible, otherwise a malfunction 
may occur. Provide a solution if the cables cannot 
be separated from a noise source due to 
unavoidable circumstances. 
Major noise sources 
Remedial measures 
Electromagnetic switch, 
etc. Power line having 
waveform distortion, 
Inverter, Thyristor regulator
Insert noise filters 
between power 
terminals and 
input/output terminals. 
A CR filter is often used.
 
  5
  Use crimp style terminals 
(1)  Mount crimp style terminals for connection 
cable terminations to prevent any looseness 
or disconnection of terminals or a short-circuit 
failure between terminals. 
(2)  Use the crimp style terminals with an 
insulation sleeve to prevent an electric 
shocks. 
 
  6
  Unused terminals
 
Don’t use any unused terminals for relaying, 
otherwise the electric circuits may be damaged. 
 
 
 
Secure the connected cables so as not to 
allow them to be hooked by a person or a 
substance, otherwise the connections may 
be cut and disrupted, and may cause an 
electric shock or other accidents.
 
 
 
Terminal block 
Screw
 
diameter
 
Tightening 
torque 
Termination (unit: mm) 
Power and 
protective 
conductor 
terminals  
M4 
1.2N • m
 
Terminals other 
than described 
above 
M3.5 
0.8N • m
*Use Type O chip (on the left) whenever possible.
 
A switch and an overcurrent 
protective device 
Turn off the power supply
before starting connections. 
Kinds of terminals and termination
 
Reference 
Warning 
Warning 
                              Secure the connected 
                              cables properly.
 
Warning
8.5 or less
8 or less 3.7 or more
3.7 or more
 
8 or less
 
4.3 or more
(with an insulation sleeve)
 
t : 0.8, with an insulation sleeve 
t : 0.8, with an insulation sleeve