Rate of Change (RoC) Filter:
This parameter helps ignore erroneous readings 
from the transmitter’s surroundings.  If the user 
knows how fast the level can change in the vessel 
(i.e. inches per second) then this parameter should 
be set at a rate that the level in the vessel cannot 
suddenly jump to.  See section “7.2 Features” for 
more details.
Trim Sensor:
This menu can be used to recalibrate the 
transmitter.
There is no need to recalibrate the 
instrument simply use the “Change Range” 
menu and change the LRV and URV as desired.
If the need arises to use this menu then 
level simulation will be required at 0 and 100 
percent.
Offset:
This parameter simply adds a digital value to the 
measuring range or span it does NOT shift or move 
the measuring range.  See section “7.2 Features” for 
more details.
The span will not change and the 4/20mA 
will not shift from its original calibrated points. 
This value also affects the reading for interface 
level when available.