Operation 
Hi Flow Sampler 
 
Instruction 0055-9017 
38 
 
3.18.4  Leak Rate Measurement – Manual 1-Stage Mode 
 
Note:
  The Manual 1-Stage Mode can only be selected if operating in the 
Expanded Menu Mode. Refer to Section 3.8. 
 
1.  If not already done, place the instrument into its Manual 1-Stage 
mode. Refer to Section 3.14.2.2 Manual 1-Stage Mode
 
2.  Choose an attachment that will ensure the complete capture of the 
gas leak. Connect this attachment to the end of the Hi Flow 
Sampler’s main sampling hose and position the attachment’s inlet 
over the leak source. 
 
3.  Position the inlet of the instrument’s background gas sampling hose 
opposite the leak source, such that the leak source will not 
contribute to the background measurement. 
 
4.  Begin the measurement process by selecting Start from the Main 
Screen.  
 
1 1 / 2 2 / 0 4   0 9 : 4 5 : 3 0  
L e a k ( % )           0 . 0 0
L e a k ( N o t   A v a i l . )
S p e e d ( L O <           > H I )
S a v e
S t a r t
[ 0 0 0 1 ]   ( M )
S - -
 
 
5.  The user is now asked to choose a new or different Test ID.  
 
Selecting Yes causes the Access Records menu to appear (refer to 
Section 3.15). After selecting or creating a new Test ID, start the 
measurement by pressing ESC until the Main Screen appears. 
Selecting No returns the display to the Main Screen and starts the 
measurement using the current Test ID. 
 
1 1 / 2 2 / 0 4   0 9 : 4 5 : 3 0  
A s s i g n   n e w   t e s t   I D ?
               
Y e s
                N o
[ 0 0 0 1 ]   ( M )
S - -
 
 
6.  Begin measuring until a stable leak measurement is achieved—this 
may take several minutes.  
 
If desired, the flow rate can be manually controlled by using the 
Speed function. Each press of the I/O key will lower the flow rate as 
indicated by the speed bar moving to the left.