Operation 
Hi Flow Sampler 
 
Instruction 0055-9017 
40 
3.19 Message 
Line 
 
Various messages will appear at the bottom of the LCD that describe 
what functions are currently being performed or errors that have been 
detected. The following list describes the meaning of these messages.  
 
Note that detailed explanations of the error messages can be found in 
Section 6.2 Troubleshooting Guide
 
Calibration Passed 
Calibration procedure was successful. 
 
Calibration Failed 
Error – The calibration procedure was not successful. Refer to 
Section 6.2. 
 
Saving Record 
A test record is currently being saved to memory. 
 
Memory Full 
Error – The instrument tried to save a test record with 1000 records 
already stored in memory. Refer to Section 6.2. 
 
No Records Saved 
An attempt was made to send records to a computer with no test 
records stored in memory. 
 
Flow Rate Lowered 
The flow rate has just been lowered signaling the start of 
measurement #2 during an automatic or manual leak test. Refer to 
Section 3.14. 
 
Erasing All Records 
All test records are being erased after selecting “Erase records” per 
Section 3.15.3. 
 
Check Background Pump 
Error – The background-gas pump’s flow rate has been severely 
reduced. Refer to Section 6.2. 
 
Check Leak Pump 
Error–The leak-gas pump’s flow rate has been severely reduced. Refer 
to Section 6.2. 
 
Check Background Calibration 
Error – The background-gas sensor needs calibrated. Refer to 
Section 6.2. 
 
Check Leak Calibration 
Error – The leak-gas sensor needs calibrated. Refer to Section 6.2. 
 
Check Background Offset 
Error – The background-gas sensor did not properly zero during 
startup. Refer to Section 6.2 
 
Check Leak Offset 
  Error – The leak-gas sensor did not properly zero during startup. Refer 
to Section 6.2.