3.19 Message
Line
Various messages will appear at the bottom of the LCD that describe
what functions are currently being performed or errors that have been
detected. The following list describes the meaning of these messages.
what functions are currently being performed or errors that have been
detected. The following list describes the meaning of these messages.
Note that detailed explanations of the error messages can be found in
Section 6.2 Troubleshooting Guide.
Calibration Passed
Calibration Passed
Calibration procedure was successful.
Calibration Failed
Error – The calibration procedure was not successful. Refer to
Section 6.2.
Section 6.2.
Saving Record
A test record is currently being saved to memory.
Memory Full
Error – The instrument tried to save a test record with 1000 records
already stored in memory. Refer to Section 6.2.
already stored in memory. Refer to Section 6.2.
No Records Saved
An attempt was made to send records to a computer with no test
records stored in memory.
Flow Rate Lowered
The flow rate has just been lowered signaling the start of
measurement #2 during an automatic or manual leak test. Refer to
Section 3.14.
measurement #2 during an automatic or manual leak test. Refer to
Section 3.14.
Erasing All Records
All test records are being erased after selecting “Erase records” per
Section 3.15.3.
Check Background Pump
Error – The background-gas pump’s flow rate has been severely
reduced. Refer to Section 6.2.
reduced. Refer to Section 6.2.
Check Leak Pump
Error–The leak-gas pump’s flow rate has been severely reduced. Refer
to Section 6.2.
Check Background Calibration
Error – The background-gas sensor needs calibrated. Refer to
Section 6.2.
Section 6.2.
Check Leak Calibration
Error – The leak-gas sensor needs calibrated. Refer to Section 6.2.
Check Background Offset
Error – The background-gas sensor did not properly zero during
startup. Refer to Section 6.2
Check Leak Offset
Error – The leak-gas sensor did not properly zero during startup. Refer
to Section 6.2.