Bacharach Hi Flow Sampler Manuel Du Propriétaire

Page de 72
Hi Flow Sampler 
Operation 
Instruction 0055-9017 
39
1 1 / 2 2 / 0 4   0 9 : 4 5 : 3 0  
F l o w ( l p m )       0 . 0
B a c k ( % )           0 . 0 0
L e a k ( % )           0 . 0 0
L e a k ( l p m )       0 . 0
S p e e d ( L O <           > H I )
[ 0 0 0 1 ]   ( M )
- 1 -
Speed Bar
 
 
 
7.  Once a stable leak measurement is obtained, all current readings can 
be saved in memory by highlighting Save and pressing the I/Okey.  
 
In the example below, the current measured leak concentration is 
15% gas, and the calculated leak rate is 42.5 lpm. 
 
1 1 / 2 2 / 0 4   0 9 : 4 7 : 1 5  
B a c k ( % )           0 . 8 0  
L e a k ( % )         1 5 . 0 0  
L e a k ( l p m )     4 2 . 5
S p e e d ( L O <           > H I )
S a v e
S a m p l e   T i m e - >   0 1 : 4 5
[ 0 0 0 1 ]   ( M )
- 1 -
 
 
8.  The measurement process continues until Stop is selected. 
 
1 1 / 2 2 / 0 4   0 9 : 4 9 : 0 0  
L e a k ( % )         1 5 . 0 0  
L e a k ( l p m )     4 2 . 5
S p e e d ( L O <           > H I )
S a v e
S t o p
S a v e
S a m p l e   T i m e - >   0 3 : 3 0
[ 0 0 0 1 ]   ( M )
- 1 -
 
 
9.  Measurements that were saved can be downloaded to a personal 
computer at a later time for analysis. Refer to Section 3.16 Importing 
Saved Data into a Spreadsheet