Bacharach Hi Flow Sampler Manuel Du Propriétaire

Page de 72
Calibration 
Hi Flow Sampler 
 
Instruction 0055-9017 
46 
5.  Apply the appropriate level of calibration gas from the calibration 
equipment to the instrument’s GAS or BACKGROUND inlet port. (In 
the example above, apply 2.5% CH
4
 to the BACKGROUND port.) Then 
press the I/Okey to start the calibration process. Observe that the 
gas-sampling-pump motors start and the calibration screen appears. 
 
1 1 / 2 2 / 0 4   1 0 : 4 5 : 3 0
B a c k ( % )           0 . 0 0
A p p l ( % )           2 . 5 0
[ 0 0 0 5 ]   ( M )
S - -
  ,         =   A d j .   A p p l i e d
E N T E R   =   C a l i b r a t e
E S C       =   E x i t
 
 
6.  If necessary, use the  keys to adjust the Appl(%) reading to match 
the gas concentration stamped on the calibration gas cylinder. 
 
7.  Wait until the measured gas reading stabilizes, and then press the 
I/O
key to calibrate the actual gas reading to that of the applied 
reading. The message “Calibration Passed!” will appear at the bottom 
of the screen if the calibration was successful. If, however, the 
calibration was not successful, the message “Calibration Failed!” will 
appear. Refer to Section 6 Troubleshooting for possible causes and 
remedies for this failure. 
 
1 1 / 2 2 / 0 4   1 1 : 4 5 : 3 0
B a c k ( % )           2 . 5 0
A p p l ( % )           2 . 5 0
[ 0 0 0 5 ]   ( M )
S - -
  ,         =   A d j .   A p p l i e d
E N T E R   =   C a l i b r a t e
E S C       =   E x i t
C a l i b r a t i o n   P a s s e d !
 
 
8.  Disconnect the gas hose, and then wait until the measured gas 
reading falls to zero percent.  
 
9.  Press the ESC key to return to the Calibration Menu Screen.  
 
10. Repeat this procedure as necessary to calibrate both sensors at 2.5% 
and 100% CH
4