Hi Flow Sampler 
Calibration 
Instruction 0055-9017 
47
4.5 
Calibration Amplifier Gain Check 
 
Use this procedure to display the gain values of the calibration amplifier 
circuits in hexadecimal. This procedure is useful to service technicians 
for troubleshooting purposes. 
 
1.  From the Main Screen select Menu > Calibration > Calibration 
Gains to display the Calibration Gains Screen. 
 
1 1 / 2 2 / 0 4   0 9 : 4 5 : 3 0
B a c k   L o w     # # # #
B a c k   H i g h   # # # #
L e a k   L o w     # # # #
L e a k   H i g h   # # # #
[ 0 0 0 1 ]   ( M )
S - -
 
 
Where: #### is the gain value in hexadecimal. 
 
2.  Press ESC three times to return to the Main Screen. 
 
 
4.6 Erase 
Calibration 
Procedure 
 
The following procedure resets the gain of each calibration amplifier 
circuit to the factory default hexadecimal value of “FFFF”. 
 
1.  From the Main Screen select Menu > Calibration > Erase 
Calibration to display the Erase Calibration Screen. 
 
1 1 / 2 2 / 0 4   0 9 : 4 5 : 3 0
E r a s e   c a l i b r a t i o n ?
             
Y e s
              N o
[ 0 0 0 1 ]   ( M )
S - -
 
 
2.  Select Yes to erase the calibration values. Observe that the message 
“Erasing Calibration” appears at the bottom of the screen. Select No 
to abort this procedure and return to the Calibration Menu Screen. 
 
3.  Press ESC twice to return to the Main Screen.