10.4  Calibration check for probe module 
FastFacts 10.4: Calibration check for probe module 
The calibration curve for the probe module is a comparison of the response of the 
probe microphone (with tube) to the response of the REM reference microphone. 
Any change in the curve indicates a change in one of these microphones (or the 
probe tube). See 
10.3: Calibration of REM probe microphone
. As part of a regular 
calibration schedule, or to investigate changing calibration curves, the REM 
reference microphone may be checked against a high quality sound level meter. 
1)  Press <Tests>, then highlight and <PICK> REM [Manual control]. 
2)  Highlight & <PICK> [Level], then highlight & <PICK> 70 from the drop-down list. 
3)  Highlight and <PICK> [Frequency], then highlight and <PICK> 2000 Hz from the 
drop-down list. 
4)  Highlight, then <PICK> [Start test] (which 
changes to [Stop test]). 
5)  Hold the probe module about 18” (0.5 m) in 
front of the REM speakers. 
6)  Hold the microphone of a calibrated sound 
level meter next to the REM reference 
microphone as shown. Set the sound level 
meter to C weighting and slow. 
7)  The sound level meter should read 70 ±2 dB. 
8)  Repeat with the frequency at 250 Hz. 
9)  To stop the measurement, highlight and 
<PICK> [Stop test]. 
The probe module should be replaced if it 
fails this test