10.3  Calibration of REM probe microphone 
FastFacts 10.3: Calibration of REM probe microphone 
Ensure the probe microphone assembly to be calibrated is plugged into the 
appropriate jack. See 
2.3: Microphone connection
 and 
10.2: REM calibration facts
 
1)  Open the test chamber and elevate the 
speaker to its vertical position. 
2)  Press <Tests>, then highlight & <PICK> 
[Calibration] from the REM column. 
3)  Press the enlarged end of a probe tube as 
far as it will go, into the recessed opening 
at the top of the probe module. 
4)  Position the open end of the probe tube in 
front of the reference microphone inlet 
and press it between the posts as shown. 
5)  Hold the probe module 5 – 6 inches (12.5 - 15 cm) away from and directly in 
front of the loudspeaker. Orientation of the probe module is not critical. 
6)  Press <PICK> to calibrate. The resulting curve should be similar in shape to 
that shown but may be shifted ± 5 dB. 
7)  To change the calibration interval, highlight & <PICK> [Daily] or [Weekly].