HP (Hewlett-Packard) RM500SL Manuel D’Utilisation

Page de 112
0611 
RM500SL User’s Guide Version 2.8  
Page 58 
13  Sensory Loss Simulator 
13.1  Sensory loss simulator description 
FastFacts 13.1: Sensory loss simulator description 
13.2  Sensory loss simulator operation 
FastFacts 13.2: Sensory loss simulator operation 
A pure conductive loss attenuates loud and soft sounds equally, and is easily 
simulated by just turning down the volume or inserting an earplug. Losses of a 
sensory nature are more complex and more difficult to simulate. 
The most common sensory loss is caused by outer hair cell damage and results in 
loss of audibility for soft sounds but near-normal loudness for loud sounds. Inner 
hair cell damage elevates threshold and reduces loudness for both soft and loud 
sounds, possibly somewhat more for soft sounds. 
This sensory loss simulator (SLS) is intended to allow normal-hearing listeners to 
hear sounds as if they had the elevated threshold and altered loudness perception 
caused by outer and inner hair cell damage. It is based on the cochlear hearing 
loss model described in Moore & Glasberg (2004). 
In this simulation, losses less than 58 dB SPL are considered to be due entirely to 
outer hair cell damage. Any loss greater than this is considered inner hair cell loss. 
This SLS does not simulate broadening of the auditory filters or other distortions 
that may accompany cochlear hearing loss. 
For comparison purposes, a simulation of a purely conductive loss is also provided. 
To properly experience the simulation, listeners should be within 1m of the REM 
speaker and in a quiet room. 
Listeners should be within 1m of the REM speaker and in a quiet room. 
1)  Press <Tests>, then highlight and <PICK> [Sensory loss simulator]. 
2)  If an audiogram has previously been entered in Speechmap or Insertion gain, 
go to step 4. Otherwise, highlight and <PICK> [Audiometry]. 
3)  Enter HL threshold values using a mouse or the keypad. Left click the mouse 
on the audiogram form to enter a point. Click again to delete the point. On the 
keypad use the arrow keys to change frequency and level and press <PICK> to 
enter or delete a threshold point. 
4)  Press <Continue> when all points have been entered. See 
16.4: Audiometric 
data entry
 or 
18.2: Assessment data entry
 for more details. 
5)  Highlight and <PICK> [Start test]. Sound, processed to simulate the entered 
hearing loss, will be presented via the REM speaker. 
6)  Highlight and <PICK> [Stimulus], then highlight and <PICK> a sound to play. 
7)  Highlight and <PICK> [Hearing], then highlight and <PICK> from [Normal, ½ 
Entered loss, Entered loss]. 
8)  Highlight and <PICK> [Type of loss] to toggle between [Sensorineural] and 
[Conductive] hearing loss simulations. 
9)  Highlight and <PICK> [Cancel] to stop the simulation.