HP (Hewlett-Packard) RM500SL Manuel D’Utilisation

Page de 112
0611 
RM500SL User’s Guide Version 2.8  
Page 60 
15  RECD (real-ear to coupler difference) 
15.1  Section overview 
FastFacts 15.1: RECD (real-ear to coupler difference) 
15.2  RECD facts 
FastFacts 15.2: RECD facts 
Individual RECD values are derived by comparing the real-ear SPL for a given test 
signal to the 2cc coupler SPL for the same test signal. These RECD values are 
used to: (1) convert HL threshold and UCL obtained using insert phones to SPL 
values, (2) simulate real-ear measurements from 2cc coupler measures (S-REM); 
and (3) derive 2cc coupler targets from real-ear aided targets.  
If you use insert ear phones
 for audiometry OR 
if you verify using S-REM
, you 
should 
measure RECD
 (preferred) or use age-related average RECD. 
If you use headphones or sound field
 for audiometry AND you 
use REM
 for 
verification, 
RECD is not applied in any way
 and need not be measured. 
If you use 2cc coupler targets, you should use measured or age-related RECD. 
Insert phone audiometry with measured RECD and REM verification provides the 
best fitting accuracy, with S-REM verification being slightly less accurate. Insert 
phone HL data with age-related average RECD is next best. 
This software provides average RECDs for 1 - 60 months, 6 - 10 years of age and 
Adult. A sampling of age-related foam-tip HA-2 RECD values is shown below. 
Age/Freq 
250 
500 
750 
1000 
1500 
2000 
3000 
4000 
6000 
1 month 
12 
15 
15 
16 
20 
23 
12 months 
10 
10 
11 
11 
15 
17 
24 months 
10 
10 
14 
15 
36 months 
13 
14 
60 months 
13 
13 
8 yrs - Adult 
13 
13 
Values are from DSL 5.0 but differ from DSL in that values for 119 months have 
been used for ages >119 months.  
DSL 5.0 RECDs differ from DSL 4.1 RECDs.
 
The RECD measurement procedure may be accessed by selecting RECD from the 
REM tests list or from the Audiometry poster within Speechmap. An optional 
RECD transducer
 is required for RECD measurements. 
This section covers the following 
topics:
  
15.2: RECD facts 
15.3: RECD coupler response 
15.4: RECD real-ear response 
15.5: RECD results