Leica M205 FA Manuel D’Utilisation

Page de 152
Leica M-series 
User Manual 
52
Under  certain  circumstances,  the  light  source 
must  not  be  moved  along  with  the  optics 
carrier. A typical example is multifocus images 
in  which  the  Z-stack  changes  while  the  angle 
of incidence of the light must remain the same. 
For such purposes, the Leica LED5000 MCI™ is 
directly fastened to the column.
Installation on the column
1.  Pull  the  retaining  stirrup  out  of  the  Leica 
LED5000 MCI™.
2.  Screw the retaining stirrup into the column 
at the notch using the single screw. If you 
tighten  the  screw,  the  retaining  stirrup 
is  automatically  moved  into  the  correct 
 position.
3.  Push  the  Leica  LED5000  MCI™  onto  the 
retaining stirrup.
Leica LED5000 MCI™: Alternative Assembly