The  ring  illuminator,  the  Leica  LED5000  RL 
("Ring Light"), is installed on the objective using 
a single screw. It has been optimized for a work-
ing distance between 60 and 70 mm.
Constraints
The Leica LED5000 RL can be used in conjunc-
tion with the planapochromat 1× and planapo-
chromat  0.63×  objectives.  With  all  other 
objectives, the working distance is too low for 
adequate illumination.
The  ring  illuminator  cannot  be  used  together 
with the objective nosepiece.
Assembly
1.  Connect the CAN-bus cable to the ring illu-
minator. The flat part of the plug must be 
facing upwards.
2.  Push the ring illuminator over the objective 
as far as it will go and screw it into place.
3.  Plug the other end of the cable into one of 
the two "CTL2" sockets on the column.
For  optimum  accessibility  of  the  specimen, 
the  ring  illuminator  should  be  installed  with 
the  cable  facing  backwards.  However,  it  is 
also possible to turn the ring illuminator side-
ways,  for  example  if  simultaneously  using  the 
Leica LED5000 MCI™ system illumination. In this 
case, the ring illuminator cannot be connected 
directly to the Leica LED5000 MCI™.
Leica LED5000 RL