5.9.4. OPTIC TEST 
The optic test function tests the response of the PMT sensor by turning on an LED 
located in the cooling block of the PMT (refer to Figure 13-18). The analyzer uses the 
light emitted from the LED to test its photo-electronic subsystem, including the PMT 
and the current to voltage converter on the pre-amplifier board. To ensure that the 
analyzer measures only the light coming from the LED, the analyzer should be supplied 
with zero air. The optic test should produce a PMT signal of about 2000±1000 mV. 
 
Access the Optic Test from the DIAG Menu (refer to Figure 5-16), then press: 
 
 
 
DIAG  
SIGNAL I / O 
 
 PREV NEXT 
JUMP 
 
         
 ENTR  
 
EXIT
 
DIAG OPTIC
        RANGE = 500.000 PPB      SO2=XXX.X 
 
<TST   TST>
 
 
 
 
    EXIT
 
Press NEXT until… 
 
 
 
DIAG  
OPTIC TEST
 
 PREV NEXT 
 
 
                      ENTR
   EXIT
 
Press TST until… 
DIAG ELEC
       
PMT = 2751  MV 
         
SO2=XXX.X  
 
<TST   TST>
 
 
 
 
   
 EXIT
 
While the optic test is 
activated,  PMT should be 
2000 mV ± 1000 mV 
 
Figure 5-30: 
DIAG – Optic Test 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
This is a coarse test for functionality and not an accurate calibration tool. 
The resulting PMT signal can vary significantly over time and also 
changes with low-level calibration. 
06807C DCN6650