Teledyne T100 Manuale Utente

Pagina di 422
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
SETUP Menu 
 
 
127 
5.9.6. LAMP CALIBRATION 
An important factor in accurately determining SO
2
 concentration is the amount of UV 
light available to transform the SO
2
 into SO
2
* (refer to Section 13.1.1).  The T100 
compensates for variations in the intensity of the available UV light by adjusting the SO
2
 
concentration calculation using a ratio (LAMP RATIO)that results from dividing the 
current UV lamp (UV LAMP) intensity by a value stored in the CPU’s memory 
(LAMP_CAL).  Both LAMP Ration and UV Lamp are test functions viewable from the 
instruments front panel.   
To cause the analyzer to measure and record a value for LAMP_CAL, access the Signal 
I/O from the DIAG Menu (refer to Figure 5-16), then press: 
 
 
 
 
 
DIAG  
SIGNAL I / O
 
NEXT 
      
          
ENTR 
  
EXIT
 
Repeat Pressing NEXT until . .
 .
 
DIAG                
LAMP CALIBRATION
        
 
  PREV  NEXT
 
      
                  ENTR  
 EXIT 
DIAG FCAL    
LAMP CAL VALUE:4262.4 mV
 
    4        2       6        2        .4                 ENTR  EXIT
 
ENTR accepts the 
new value  
EXIT 
ignores the new 
value  
The value displayed is the 
current output of the UV 
source reference detector 
 
Figure 5-32: 
DIAG – Lamp Calibration 
 
 
06807C DCN6650