5.9.5. ELECTRICAL TEST 
The electrical test function creates a current, which substitutes the PMT signal, and 
feeds it into the preamplifier board. This signal is generated by circuitry on the pre-
amplifier board itself and tests the filtering and amplification functions of that assembly 
along with the A/D converter on the motherboard. It does not test the PMT itself. The 
electrical test should produce a PMT signal of about 2000 ±1000 mV.  
 
Access the Electrical Test from the DIAG Menu (refer to Figure 5-16), then press: 
 
 
 
DIAG  
SIGNAL I / O  
 
 PREV 
NEXT 
JUMP 
 
        
  ENTR
 
  
EXIT
 
DIAG ELEC 
        RANGE = 500.0 PPB         SO2 =XXX.X 
  
<TST   TST>
 
 
 
 
    EXIT
 
Press NEXT until… 
 
 
 
DIAG  
ELECTRICAL TEST  
 
 PREV NEXT 
 
 
                      ENTR
   EXIT
 
Press TST until… 
DIAG ELEC
       
PMT = 1732  MV 
        
SO2=X.XXX 
  
<TST   TST>
 
 
 
 
   
 EXIT
 
While the electrical test is 
activated,  PMT should equal: 
2000 mV ± 1000 mV 
 
Figure 5-31: 
DIAG – Electrical Test 
 
 
06807C DCN6650