SETUP Menu 
 
 
129 
5.9.8. FLOW CALIBRATION 
The flow calibration allows the user to adjust the values of the sample flow rates as they 
are displayed on the front panel and reported through COM ports to match the actual 
flow rate measured at the sample inlet. This does not change the hardware measurement 
of the flow sensors, only the software calculated values.  
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the 
SAMPLE inlet (refer to Section 11 for more details).  
Once the flow meter is attached and is measuring actual gas flow, access the Signal I/O 
from the DIAG Menu (refer to Figure 5-16), then press:  
 
 
 
 
 
DIAG  
SIGNAL I / O
 
NEXT 
      
          
ENTR 
  
EXIT
 
Repeat Pressing NEXT until . .
 .
 
DIAG                
FLOW CALIBRATION
        
 
 
  PREV  NEXT
 
      
                  ENTR  
 EXIT 
DIAG FCAL       
ACTUAL FLOW: 607 CC / M 
 
    0        6      0        7                           ENTR  EXIT
 
ENTR accepts the 
new value  
EXIT 
ignores the new 
value  
Adjust these values until the 
displayed flow rate equals the 
flow rate being measured by the 
independent flow meter. 
 
Figure 5-34: 
DIAG – Flow Calibration 
 