5.9.7. PRESSURE CALIBRATION 
A sensor at the exit of the sample chamber continuously measures the pressure of the 
sample gas.  This data is used to compensate the final SO
2
 concentration calculation for 
changes in atmospheric pressure when the instrument’s TPC feature is turned on (refer 
to Section 10.7.3) and is stored in the CPU’s memory as the test function PRES (also 
viewable via the front panel). 
Ensure to use a barometer that measures actual barometric pressure. 
To cause the analyzer to measure and record a value for PRES, access the Signal I/O 
from the DIAG Menu (refer to Figure 5-16), then press: 
 
 
 
 
 
 
 
 
 
DIAG  
SIGNAL I / O  
 
NEXT 
      
          
ENTR 
  
EXIT
 
Repeat Pressing NEXT until . .
 .
 
DIAG               
PRESSURE CALIBRATION
        
 
 
  PREV  NEXT
 
      
                  ENTR  
 EXIT 
DIAG PCAL   
ACTUAL PRES :27.20 IN-HG-A 
 
     2       7       .2       0                           ENTR  EXIT
 
ENTR accepts the 
new value  
EXIT 
ignores the new 
value  
Adjust these values until the 
displayed pressure equals the 
pressure measured by the 
independent pressure meter. 
 
Figure 5-33: 
DIAG – Pressure Calibration 
 