5.4.5. DILUTION RATIO (OPTION) 
This feature is a optional software utility that allows the user to compensate for any 
dilution of the sample gas that may occur before it enters the sample inlet. Typically this 
occurs in continuous emission monitoring (CEM) applications where the sampling 
method used to remove the gas from the stack dilutes it.   
Using the dilution ratio option is a 4-step process: 
1.  Select the appropriate units of measure (see Section 5.4.4). 
2.  Select the reporting range mode and set the reporting range upper limit (see Section 
  Ensure that the upper span limit entered for the reporting range is the maximum 
expected concentration of the UNDILUTED gas.  
3.  Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1 
part of sample gas): 
 
 
4.  Calibrate the analyzer.  
  Make sure that the calibration span gas is either supplied through the same dilution 
system as the sample gas or has an appropriately lower actual concentration.  
EXAMPLE:  If the reporting range limit  is set for 100 ppm and the dilution ratio 
of the sample gas is 20 gain, either: 
  a span gas with the concentration of 100 ppm can be used if the span gas passes 
through the same dilution steps as the sample gas, or; 
   a 5 ppm span gas must be used if the span gas IS NOT routed through the dilution 
system.  