Setup Menu 
109 
 
5.5. SETUP  PASS: PASSWORD PROTECTION 
The menu system provides password protection of the calibration and setup functions to 
prevent unauthorized adjustments.  When the passwords have been enabled in the PASS 
menu item, the system will prompt the user for a password anytime a password-
protected function (e.g., SETUP) is selected. This allows normal operation of the 
instrument, but requires the password (101) to access to the menus under SETUP. When 
PASSWORD is disabled (SETUP>OFF), any operator can enter the Primary Setup 
(SETUP) and Secondary Setup (SETUP>MORE) menus. Whether PASSWORD is 
enabled or disabled, a password (default 818) is required to enter the VARS or DIAG 
menus in the SETUP>MORE menu. 
Table 5-2: 
Password Levels 
PASSWORD
LEVEL 
MENU ACCESS ALLOWED 
Null (000) 
Operation 
All functions of the main menu (top level, or Primary,  menu) 
101 
Configuration/Maintenance 
Access to Primary and Secondary SETUP Menus when PASSWORD is 
enabled 
818 
Configuration/Maintenance  Access to Secondary SETUP Submenus VARS and DIAG whether 
PASSWORD is enabled or disabled. 
 
To enable or disable passwords, press: 
 
If the password feature is enabled, then when entering either Calibration or Setup Mode, 
the default password displayed will be 000, and the new password must be input. 
Example: If all passwords are enabled, the following control button sequence would be 
required to enter the SETUP menu: 