Teledyne t320u Manuale Utente

Pagina di 426
Teledyne API – Model T300/T300M CO Analyzer 
Calibration Procedures 
209 
9.3.3. MANUAL CALIBRATION USING VALVE OPTIONS 
The following section describes the basic method for manually calibrating the 
T300/T300M Analyzer. 
If the analyzer’s reporting range is set for the DUAL or  AUTO range modes, a step will 
appear for selecting which range is to be calibrated (LOW or HIGH).  
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Each of these two ranges MUST be calibrated separately. 
 
9.3.3.1. SETTING THE EXPECTED SPAN GAS CONCENTRATION 
Note 
When setting expected concentration values, consider impurities in your 
span gas. 
The expected CO span gas concentration should be 80% of the reporting range of the 
instrument (see Section 5.4.1).  The default factory setting is 40 ppm.   
To set the span gas concentration, press: 
06864B DCN6314