Calibration Procedures 
Teledyne API – Model T300/T300M CO Analyzer 
210 
 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
For this Initial Calibration it is important to independently verify the 
PRECISE CO Concentration Value of the SPAN gas. 
If the source of the Span Gas is from a Calibrated Bottle, use the exact 
concentration value printed on the bottle. 
9.3.3.2. ZERO/SPAN POINT CALIBRATION PROCEDURE 
The zero and cal operations are initiated directly and independently with dedicated 
buttons (CALZ & CALS). 