Calibration Procedures 
Teledyne API – Model T300/T300M CO Analyzer 
212 
9.3.3.3. USE OF ZERO/SPAN VALVE WITH REMOTE CONTACT CLOSURE 
Contact closures for controlling calibration and calibration checks are located on the rear 
panel CONTROL IN connector.  Instructions for setup and use of these contacts can be 
found in Section 3.3.1.6. 
When the appropriate contacts are closed for at least 5 seconds, the instrument switches 
into zero, or span calibration mode and any internal zero/span valves installed will be 
automatically switched to the appropriate configuration.   
  The remote calibration contact closures may be activated in any order.   
  It is recommended that contact closures remain closed for at least 10 minutes to 
establish a reliable reading. 
  The instrument will stay in the selected mode for as long as the contacts remain 
closed. 
If contact closures are being used in conjunction with the analyzer’s AutoCal (see 
Section  9.4) feature and the AutoCal attribute CALIBRATE” is enabled, the 
T300/T300M will not recalibrate the analyzer until the contact is opened.  At this point, 
the new calibration values will be recorded before the instrument returns to Sample 
Mode
.   
If the AutoCal attribute “CALIBRATE” is disabled, the instrument will return to 
Sample Mode, leaving the instrument’s internal calibration variables unchanged. 
9.4. AUTOMATIC ZERO/SPAN CAL/CHECK (AUTOCAL) 
The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve 
options by using the T300/T300M Analyzer’s internal time of day clock.  AutoCal 
operates by executing SEQUENCES programmed by the user to initiate the various 
calibration modes of the analyzer and open and close valves appropriately.  It is possible 
to program and run up to three separate sequences (SEQ1,  SEQ2 and SEQ3).  Each 
sequence can operate in one of three modes, or be disabled. 
Table 9-2: 
AUTOCAL Modes 
MODE NAME 
ACTION 
DISABLED 
Disables the Sequence. 
ZERO 
Causes the Sequence to perform a Zero calibration/check. 
ZERO-SPAN 
Causes the Sequence to perform a Zero point calibration/check followed by a 
Span point calibration/check. 
SPAN 
Causes the Sequence to perform a Span concentration calibration/check only. 