Teledyne API – Model T300/T300M CO Analyzer 
Calibration Procedures 
219 
9.6.  CALIBRATION OF THE T300/T300M’S ELECTRONIC 
SUBSYSTEMS 
9.6.1. DARK CALIBRATION TEST 
The dark calibration test interrupts the signal path between the IR photo-detector and the 
remainder of the sync/demod board circuitry.  This allows the instrument to compensate 
for any voltage levels inherent in the sync/demod circuitry that might effect the 
calculation of CO concentration.   
Performing this calibration returns two offset voltages, one for CO MEAS and one for 
CO REF that are automatically added to the CPU’s calculation routine.  The two offset 
voltages from the last calibration procedure may be reviewed by the user via the front 
panel display. (See also Section 5.9.5). 
To activate the dark calibration procedure or review the results of a previous calibration, 
press: 