Calibration Procedures 
Teledyne API – Model T300/T300M CO Analyzer 
220 
Offset for CO REF signal
SAMPLE
RANGE=50.0 PPM          CO= XX.XX
<TST
TST> CAL
SETUP
SETUP X.X         
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X       
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
DIAG                        
SIGNAL I/O
PREV
NEXT
ENTR
EXIT
SETUP X.X     
ENTER PASSWORD
8
1
8
ENTR
EXIT
Continue pressing NEXT until ...
DIAG OPTIC            
DARK CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG DARK   
CO DARK CALIBRATION
VIEW
CAL
EXIT
DIAG DARK   
REF DARK OFFSET: 0.0mV
EXIT
Offset for CO MEAS signal
DIAG DARK   
MEAS DARK OFFSET: 0.0mV
EXIT
DIAG DARK   
DARK CAL 1% COMPLETE
EXIT
DIAG DARK   
DARK CALIBRATION ABORTED
EXIT
Calibration runs automatically
 
9.6.2. PRESSURE CALIBRATION 
A sensor at the sample chamber outlet continuously measures the pressure of the sample 
gas.  These data are used to compensate the final CO concentration calculation for 
changes in atmospheric pressure and is stored in the CPU’s memory as the test function 
PRES (also viewable via the front panel).  See also Section 5.9.6. 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
This calibration must be performed when the pressure of the sample gas 
is equal to ambient atmospheric pressure. Before performing the 
following pressure calibration procedure, disconnect the sample gas 
pump and sample gas-line vent from the rear panel sample gas inlet. 
06864B DCN6314