EPA Calibration Protocol 
Teledyne API – Model T300/T300M CO Analyzer 
240 
If the linearity error is excessive and cannot be attributed to outside causes, check the 
T300 system for: 
  Sample pressure higher than ambient – pressurized sample gas  
 Leaks 
 Correct 
flow 
  Miscalibrated span gas tanks or bad zero gas 
  Miscalibrated sample pressure transducer 
  Failed IR detector, GFC Wheel or Sync/Demod Board 
  Contaminated optical bench or sample lines 
10.6. REFERENCES 
Quality Assurance Handbook for Air Pollution Measurement Systems Volume II: Part 
1 - Ambient Air Quality Monitoring Program Quality System Development - EPA-
454/R-98-004 - August 1998.  United States Environmental Protection Agency - 
Office of Air Quality Planning and Standards  
CFR Title 40: Protection of Environment - PART 53—AMBIENT AIR MONITORING 
REFERENCE AND EQUIVALENT METHODS: 
 - 53.20   General provisions. 
 - 53.23   Test procedures. 
CFR Title 40: Protection of Environment - PART 50—NATIONAL PRIMARY AND 
SECONDARY AMBIENT AIR QUALITY STANDARDS: Appendix C to Part 50—
Measurement Principle and Calibration Procedure for the Measurement of Carbon 
Monoxide in the Atmosphere (Non-Dispersive Infrared Photometry) 
4
 Quality Assurance Handbook for Air Pollution Measurement Systems - Volume II, 
Ambient Air Specific Methods, EPA-600/4-77-027a, 1977. 
5
  CFR Title 40: Protection of Environment - AMBIENT AIR QUALITY SURVEILLANCE  
 
06864B DCN6314