year.  Each agency must audit 25% of the reference or equivalent analyzers each quarter.  
If an agency operates less than four reference or equivalent analyzers, it must randomly 
select analyzers for reauditing so that one analyzer will be audited each calendar quarter 
and each analyzer will be audited at least once a year. 
Appendix B, 40 CFR 58
5
 requires that each Prevention of Significant Deterioration 
(PSD) reference or equivalent analyzer be audited at least once a sampling quarter.  
Results of these audits are used to estimate the accuracy of ambient air data. 
 
10.4.2. DATA REDUCTION AUDIT 
A data reduction audit involves transcribing analyzer data and determining if the 
collected data is within the control limits, generally 
2 ppm between the analyzer 
response and the audit value.  The resulting values are recorded on the SAROAD form.  
If data exceeds 
2 ppm, check all of the remaining data in the 2-week period. 
 
10.4.3. SYSTEM AUDIT/VALIDATION 
A system audit is an on-site inspection and review of the quality assurance activities 
used for the total measurement system (sample collection, sample analysis, data 
processing, etc.); it is an appraisal of system quality. 
Conduct a system audit at the startup of a new monitoring system and periodically (as 
appropriate) as significant changes in system operations occur. 
 
10.5. DYNAMIC MULTIPOINT CALIBRATION PROCEDURE 
10.5.1. LINEARITY TEST 
In order to record the instrument’s performance at a predetermined sensitivity and to 
derive a calibration relationship, a minimum of three reference points and one zero point 
uniformly spaced covering 0 to 90 percent of the operating range are recommended to 
define this relationship. 
The analyzer's recorded response is compared with the known concentration to derive 
the calibration relationship.  
To perform a precision check during the instrument set up, the sources of zero air and 
sample gas should conform to those described in Section 9.1.1.3. 
Follow the procedures described in Section 9.3 for calibrating the zero points.   