Teledyne API – Model T300/T300M CO Analyzer 
Troubleshooting and Service 
293
 
12.6.3.2. ADJUSTING THE SYNC/DEMODULATOR, CIRCUIT GAIN 
To adjust the sync/demodulator circuit gain: 
1.  Make sure that the analyzer is turned on and warmed up. 
2.  Set the analyzer display to show the STABIL or CO STB test function. 
3.  Apply Zero Air to Sample Inlet of the analyzer. 
4.  Wait until the stability reading falls below 1.0 ppm. 
5.  Change the analyzer display to show the CO MEAS. 
6.  Remove the Sync/Demod Housing 
  Remove the two mounting screws. 
  Carefully lift the housing to reveal the sync/demod PCA. 
Sync/Demod 
PCA Housing 
Housing Mounting 
Screws 
Optical Bench 
 
Figure 12-19:  Location of Sync/Demod Housing Mounting Screws 
 
7.  Adjust  potentiometer VR1 until CO MEAS reads 4500 mV ± 300 mV 
 
VR1
Adjustment Made Here 
 
Figure 12-20:  Location of Sync/Demod Gain Potentiometer 