Theory of Operation 
299
 
 
IR 
Source 
IR Beam 
Sample Chamber 
Band-Pass Filter 
Photo-Detector 
 
Figure 13-1: 
Measurement Fundamentals 
Upon exiting the sample cell, the beam shines through a band-pass filter that allows only 
light at a wavelength of 4.7 µm to pass.  Finally, the beam strikes a solid-state photo-
detector that converts the light signal into a modulated voltage signal representing the 
attenuated intensity of the beam.   
 
13.2.1. GAS FILTER CORRELATION 
Unfortunately, water vapor absorbs light at 4.7 µm too.  To overcome the interfering 
effects of water vapor the T300/T300M adds another component to the IR light path 
called a Gas Filter Correlation (GFC) Wheel. 
Measurement Cell
(Pure N
2
)
 
Reference Cell 
(N
2
 with CO)
 
 
Figure 13-2: 
GFC Wheel 
13.2.1.1. THE GFC WHEEL 
A GFC Wheel is a metallic wheel into which two chambers are carved.  The chambers 
are sealed on both sides with material transparent to 4.7 µm IR radiation creating two 
airtight cavities.  Each cavity is mainly filled with composed gases.  One cell is filled 
with pure N
2
 (the measurement cell)
.  
The other is filled with a combination of N
2
 and a 
high concentration of CO (the reference cell). 
06864B DCN6314