Theory of Operation 
Teledyne API – Model T300/T300M CO Analyzer 
300 
 
IR 
Source 
Photo-Detector 
M
 
 
R
 
GFC Wheel 
IR unaffected by N
2
 in Measurement Cell 
IR is affected by CO in Reference Cell 
Δ H 
 
Figure 13-3: 
Measurement Fundamentals with GFC Wheel 
 
As the GFC Wheel spins, the IR light alternately passes through the two cavities.  When 
the beam is exposed to the reference cell, the CO in the gas filter wheel strips the beam 
of most of the IR at 4.7μm.  When the light beam is exposed to the measurement cell, 
the N
2
 in the filter wheel does not absorb IR light.  This causes a fluctuation in the 
intensity of the IR light striking the photo-detector which results in the output of the 
detector resembling a square wave. 
13.2.1.2. THE MEASURE REFERENCE RATIO 
The T300/T300M determines the amount of CO in the sample chamber by computing 
the ratio between the peak of the measurement pulse (CO MEAS) and the peak of the 
reference pulse (CO REF).   
If no gases exist in the sample chamber that absorb light at 4.7μm, the high 
concentration of CO in the gas mixture of the reference cell will attenuate the intensity 
of the IR beam by 60% giving a M/R ratio of approximately 2.4:1.   
Adding CO to the sample chamber causes the peaks corresponding to both cells to be 
attenuated by a further percentage.  Since the intensity of the light passing through the 
measurement cell is greater, the effect of this additional attenuation is greater.  This 
causes CO MEAS to be more sensitive to the presence of CO in the sample chamber 
than  CO REF and the ratio between them (M/R) to move closer to 1:1 as the 
concentration of CO in the sample chamber increases. 