SPECIFICATIONS AND APPROVALS  
This section presents specifications for the T300/T300M analyzer and for its second gas 
sensor options, EPA equivalency designation, and compliance statements. 
2.1. SPECIFICATIONS 
Table 2-1:  T300/T300M Basic Unit Specifications 
Parameter  
Specification  
Ranges 
Min: 0-1 ppm Full scale 
Max: 0-1,000 ppm Full scale (selectable, dual ranges and auto ranging supported) 
Measurement Units 
T300: ppb, ppm, µg/m
3
, mg/m
3
 (user selectable) 
T300M: ppm, mg/m
3
 (user selectable) 
Zero Noise
1
 
T300: < 0.02 ppm RMS 
T300M: ≤ 0.1 ppm RMS 
Span Noise
1
 
T300:<0.5% of rdg RMS over 5ppm
 3
 
T300M:>0.5% of rdg RMS over 20ppm
 
Lower Detectable Limit
1
 
T300: < 0.04 ppm 
T300M: 0.2 ppm 
Zero Drift (24 hours)
 2
 
T300: < 0.1 ppm 
T300M: <0.5 ppm 
Span Drift (24 hours)
 2
 
T300: < 0.5% of reading 
T300M: 0.5ppm 
Lag Time
1
 10 
seconds 
Rise/Fall Time 
1
 
<60 seconds to 95% 
Linearity 
T300: 1% of full scale
5
;
 
T300M: 0 - 3000 ppm: 1% full scale;   
3000 - 5000 ppm: 2% full scale
 
Precision 
T300: The greater of 0.5% of reading or 0.2ppm;  
T300M: The greater of 1.0% of reading or 1ppm 
Sample Flow Rate 
800 cm
3
/min.  ±10% 
(O
2
 Sensor option adds 120 cm³/min to total flow when installed) 
AC Power 
100V-120V, 220V-240V, 50/60 Hz 
Analog Output Ranges 
All Outputs: 10V, 5V, 1V, 0.1V (selectable)  
Three outputs convertible to 4-20 mA isolated current loop. 
All Ranges with 5% under/over-range 
Analog Output Resolution 
1 part in 4096 of selected full-scale voltage 
Recorder Offset 
±10% 
Standard I/O 
1 Ethernet: 10/100Base-T 
2 RS-232 (300 – 115,200 baud)  
2 USB device ports 
8 opto-isolated digital status outputs 
6 opto-isolated digital control inputs (2 defined, 4 spare) 
4 user configurable analog outputs 
06864B DCN6314