Getting Started 
51 
Table 3-6:  Status Output Signals 
REAR PANEL 
LABEL 
STATUS 
DEFINITION 
CONDITION 
SYSTEM OK 
ON if no faults are present. 
2 CONC 
VALID 
OFF any time the HOLD OFF feature is active, such as during calibration or when 
other faults exist possibly invalidating the current concentration measurement 
(example: sample flow rate is outside of acceptable limits). 
ON if concentration measurement is valid.   
3 HIGH 
RANGE 
ON if unit is in high range of either the DUAL or AUTO range modes. 
4 ZERO 
CAL 
ON whenever the instrument’s ZERO point is being calibrated. 
5 SPAN 
CAL 
ON whenever the instrument’s SPAN point is being calibrated. 
6 DIAG 
MODE 
ON whenever the instrument is in DIAGNOSTIC mode. 
7 CO
2
 CAL
 
If this analyzer is equipped with an optional CO
2
 sensor, this Output is ON when that 
sensor is in calibration mode.  
Otherwise this output is unused. 
8 O
2
 CAL
 
If this analyzer is equipped with an optional O
2
 sensor, this Output is ON when that 
sensor is in calibration mode. 
Otherwise this output is unused. 
EMITTER BUS 
The emitters of the transistors on pins 1-8 are bussed together. 
 
SPARE 
DC POWER 
+ 5 VDC, 300 mA source maximum. 
 
 
Digital Ground 
The ground level from the analyzer’s internal DC power supplies. 
3.3.1.6. CONNECTING THE CONTROL INPUTS 
To remotely activate the zero and span calibration modes, several digital control inputs 
are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear 
panel.   
There are two methods for energizing the control inputs.  The internal +5V available 
from the pin labeled “+” is the most convenient method.  However, if full isolation is 
required, an external 5 VDC power supply should be used. 