Teledyne t320u Manuale Utente

Pagina di 426
Teledyne API – Model T300/T300M CO Analyzer 
Getting Started 
53 
3.3.1.7. CONNECTING THE CONCENTRATION ALARM RELAY (OPTION 61) 
The concentration alarm option is comprised of four (4) “dry contact” relays on the rear 
panel of the instrument.  This relay option is different from and in addition to the 
“Contact Closures” that come standard on all Teledyne API instruments.  Each relay has  
3 pins: Normally Open (NO), Common (C),  and Normally Closed (NC).   
 
Figure 3-13: 
Concentration Alarm Relay 
 
 
Alarm 1  
“System OK 2” 
Alarm 2 
“Conc 1” 
Alarm 3 
“Conc 2” 
Alarm 4 
“Range Bit” 
 
“A
LARM 
1”
 
R
ELAY 
 
Alarm 1 which is “System OK 2” (system OK 1, is the status bit) is in the energized 
state when the instrument is “OK” & there are no warnings.  If there is a warning active 
or if the instrument is put into the “DIAG” mode, Alarm 1 will change states.  This 
alarm has “reverse logic” meaning that if you put a meter across the Common & 
Normally Closed pins on the connector you will find that it is OPEN when the 
instrument is OK.  This is so that if the instrument should turn off or lose power, it will 
change states and you can record this with a data logger or other recording device. 
“A
LARM 
2”
 
R
ELAY 
&
 
“A
LARM 
3”
 
R
ELAY
 
Alarm 2 relay is associated with the “Concentration Alarm 1” set point in the software; 
Alarm 3 Relay is associated with the “Concentration Alarm 2” set point in the software.   
Alarm 2 Relay  
CO Alarm 1 = xxx PPM 
Alarm 3 Relay  
CO
2
 Alarm 2 = xxx PPM 
Alarm 2 Relay   
CO Alarm 1 = xxx PPM 
Alarm 3 Relay  
CO
2
 Alarm 2 = xxx PPM 
The Alarm 2 Relay will be turned on any time the concentration set-point is exceeded & 
will return to its normal state when the concentration value goes back below the 
concentration set-point.   
 
Even though the relay on the rear panel is a NON-Latching alarm & resets when the 
concentration goes back below the alarm set point, the warning on the front panel of the 
06864B DCN6314