Getting Started 
Teledyne API – Model T300/T300M CO Analyzer 
70 
3.3.2.7. PNEUMATIC CONNECTIONS FOR ZERO SCRUBBER/PRESSURIZED SPAN OPTION 
 
Figure 3-25: 
Pneumatic Connections – Zero Scrubber/Pressurized Span Calibration Valves (Opt 50E) 
S
AMPLE 
G
AS 
S
OURCE  
 
Attach a sample inlet line to the sample inlet port.  The SAMPLE input line should not 
be more than 2 meters long. 
  Maximum pressure of any gas at the sample inlet should not exceed 1.5 in-hg above 
ambient pressure and ideally should equal ambient atmospheric pressure. 
  In applications where the sample gas is received from a pressurized manifold, a 
vent must be placed on the sample gas before it enters the analyzer. 
C
ALIBRATION 
G
AS 
S
OURCES 
 
S
PAN 
G
AS
:
 
 
  Attach a gas line from the pressurized source of calibration gas (e.g. a bottle of 
NIST-SRM gas) to the span inlet. 
  Span gas can by generated by a T700 Dynamic Dilution Calibrator.   
Z
ERO 
A
IR
:
 
 
  Zero air is supplied internally via a zero air scrubber that draws ambient air through 
the ZERO AIR inlet. 
I
NPUT 
G
AS 
V
ENTING 
 
The zero air supply and sample gas line MUST be vented in order to ensure that the 
gases input do not exceed the maximum inlet pressure of the analyzer as well as to 
prevent back diffusion and pressure effects. These vents should be: 
  At least 0.2m long;  
  No more than 2m long and;  
  Vented outside the shelter or immediate area surrounding the instrument. 
A similar vent line should be connected to the VENT SPAN outlet on the back of the 
analyzer. 