Teledyne API – Model T300/T300M CO Analyzer 
Getting Started 
71 
E
XHAUST 
O
UTLET  
 
Attach an exhaust line to the analyzer’s EXHAUST outlet fitting.  The exhaust line 
should be: 
  PTEF tubing; minimum O.D ¼”; 
  A maximum of 10 meters long; 
  Vented outside the analyzer’s enclosure. 
3.3.2.8. PNEUMATIC LAYOUT FOR ZERO SCRUBBER/PRESSURIZED SPAN OPTION 
 
Figure 3-26: 
Internal Pneumatic Flow OPT 50E – Zero Scrubber/Pressurized Span 
 
Table 3-10:  Zero/Span Valve Operating States for Option 51E 
Mode Valve 
Condition 
Sample/Cal 
Open to SAMPLE inlet 
Zero/Span 
Open to internal ZERO AIR 
scrubber  
SAMPLE 
(Normal State) 
Shutoff Valve 
Closed 
Sample/Cal 
Open to zero/span valve 
Zero/Span 
Open to internal ZERO AIR 
scrubber 
ZERO CAL 
Shutoff Valve 
Closed 
Sample/Cal 
Open to ZERO/SPAN valve 
Zero/Span 
Open to SHUTOFF valve 
SPAN CAL 
Shutoff Valve 
Open to PRESSURE SPAN 
inlet 