6 - 14
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Technical level configuration
↓↓
System set up
↓↓
System Reset
↓↓
You can start the following functions in the menu "System Reset":
1) System Reset:
♦ 
The MLT analyzer or the platform will change to the initializing mode automatically, if
you press the ENTER key or the 
 -key in the line "System reset!":
This reboot is analogous to the "switching on" of the control module.
2) Re-initialize network:
♦ 
Change to the line "Re-initialize network!" with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Press the ENTER key or the 
 -key to start the function.
Caution: With the function "Re-initialize network!", you will delete all bindings between
analyzer modules and I/O modules and all configuration data of the SIO and DIO modules
will disappear!
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
System Reset
Are you sure?
Re-initializing will destroy all the binds!!!
Re-initialize network!
System reset!
6.1.5 System set up
System Reset