Insulet Corporation 017 Manuale Utente
127
10
Occlusion detected.
Insulin delivery stopped.
Change Pod now.
Hazard
Alarm
Alarm
Pod
and
PDM
and
PDM
Continuous tone.
Press OK to acknowledge.
Change Pod.
Check blood glucose.
Pod Error.
Insulin delivery stopped.
Change pod now.
Hazard
Alarm
Alarm
Pod
or
PDM
or
PDM
Continuous tone.
This alarm occurs when the Pod detects
an unexpected error. Press OK to
acknowledge.
an unexpected error. Press OK to
acknowledge.
Change Pod.
Check blood glucose.
PDM Error.
Remove Pod now.
Remove Pod now.
Call Customer Care:
800-591-3455.
800-591-3455.
Hazard
Alarm
Alarm
PDM
Continuous tone.
This alarm occurs if an unexpected error
is detected in the PDM. Press OK to
acknowledge.
is detected in the PDM. Press OK to
acknowledge.
Remove Pod.
Call Customer Care immediately.
Check blood glucose.
Alarm
Type
Source
Beep Pattern
Action