152
Pod Specifications
Size: 3.9 cm x 5.2 cm x 1.45 cm
Weight (without insulin): 25 grams 
Operating temperature range: 4.4°C to 40°C 
Under normal circumstances, your body temperature will keep 
the Pod well within this range.
Startup temperature: above 10°C
Storage temperature range: 0°C to 30°C
Reservoir volume (deliverable): 200 units
Cannula insertion depth: 6.5 mm
Waterproof rating: IPX8 (7.6 meters for up to 60 minutes)
Insulin concentration: U-100
Alarm type: Audible
Operating relative humidity range: 20–85%, non-condensing
Storage relative humidity range: 20–85%, non-condensing
Operating atmospheric pressure: 696 hPA to 1060 hPA
Storage atmospheric pressure: 696 hPA to 1060 hPA
Non-pyrogenic: Fluid pathway only
Type BF Medical Device: Protection from electrical shock
Flow Capability:
         Basal: Programmable by the user in 0.05 U pulses 
         up to 30.0 U per hour
     Bolus:
Rate: 1.5 units per minute
Range: 0.05 to 30.0 units 
Delivery accuracy (tested per IEC 60601-2-24):
         Basal: +/- 5% at rates > 0.05 U/hr
         Bolus: +/- 5% for all set values > 0.05 units
Bolus presets
7
Carbohydrate presets
36
Suspend
30 minutes to 2 hours
Low reservoir volume 
indicator
10 to 50 units in 5-unit increments
Default is 10.0 units
Pod expiration alert
1 to 24 hours in 1-hour increments
Default is 4 hours
Auto-off alert
1 to 24 hours or Off, in 1-hour 
increments
Default is Off
PDM lock
On or Off
Default is Off
History storage
5400 records/90 days
Language
English