Insulet Corporation 019 Manuale Utente

Pagina di 190
153
Accuracy Test Results
The following graph is designed to show flow accuracy of the 
Pod against given time periods.
Flow rate error
The following trumpet curve shows the accuracy of the delivery 
rate in relation to the observation period. The measurements 
were made using a Pod with a basal rate of 0.05 U/h at high oper-
ating temperature. The overall mean percentage flow error was 
1.40%.