Personal Diabetes Manager Specifications
Size: 6.21cm x 11.25cm x 2.5cm
Weight (with batteries): 125 grams 
Screen: 3.6cm wide x 4.8cm long; 6.1cm diagonal screen LCD
Battery: Powered by (2) AAA alkaline batteries
Battery life: Approximately 3 weeks
Operating temperature range: 4.4°C to 40°C
Storage temperature range: -29°C to 60°C
Operating relative humidity range: 10% to 90% non-
condensing
Storage relative humidity range: 10% to 90% non-condensing
Operating atmospheric pressure: 696 hPA to 1062 hPA
Storage atmospheric pressure: 703 hPA to 1062 hPA
Communication distance: At start up the 
PDM and Pod should 
be adjacent and touching, either in or out of tray to ensure proper 
communication during priming. At least 152.40 cm during 
normal operation.
Alarm type: Audible
Alert type: Audible and vibratory
Warranty: 4 Year Limited (PDM)
Blood Glucose Meter Specifications
Assay method: Coulometric electrochemical sensor
Calibration: Plasma equivalent
Hematocrit: 15% to 65%
Measurement units: mg/dL
Result range: 20 to 500 mg/dL
Sample: Whole blood, capillary
Sample size: 300 nanoliters (0.3 microliters)
Test time: Results in as fast as 7 seconds