The Programmer System can be used to:
Interrogate, adjust, and monitor the therapies being delivered by the IPG
Monitor IPG status information such as battery voltage and end of service indicators
Monitor IPG status information such as battery voltage and end of service indicators
Navigating
Use the pointing device to navigate the Programmer Software. In this section, the word “click” indicates
depressing the left pointing device button to perform the required action.
depressing the left pointing device button to perform the required action.
User Login
After powering up the programmer you will be requested to login. Click the CVRx User and enter the
password.
password.
Starting the Application
Double-click the “CVRx” icon labeled “CVRx Launcher” to start the software application.
NOTE: When the software is started, it automatically looks for the Programmer Interface. If the
software cannot find the Programmer Interface, a warning message will be displayed. Plug the
Programmer Interface into the computer USB port before continuing.
software cannot find the Programmer Interface, a warning message will be displayed. Plug the
Programmer Interface into the computer USB port before continuing.
Connection Dialog
When the Connection Dialog window appears, verify that the system time, date, and time zone on the
programmer is correct. If not, click the “Set…” button to correct the time.
programmer is correct. If not, click the “Set…” button to correct the time.
The Discovery status should indicate “In progress…” While Discovery is in progress, all Neo and Neo
Legacy IPG devices within telemetry range will be displayed in the list of IPGs discovered.
Legacy IPG devices within telemetry range will be displayed in the list of IPGs discovered.
In order to establish communication, first select the device by clicking on the row containing the desired
patient identification and serial number information. This will highlight the row. Ensure that the patient
identification and serial number of the selected IPG matches that of the device implanted in the patient
being treated. If an incorrect device is selected, a telemetry link will be established with an IPG that is not
the one implanted in the patient being treated.
patient identification and serial number information. This will highlight the row. Ensure that the patient
identification and serial number of the selected IPG matches that of the device implanted in the patient
being treated. If an incorrect device is selected, a telemetry link will be established with an IPG that is not
the one implanted in the patient being treated.
Note: If the desired IPG does not appear on the list, move the Programmer Interface closer to the patient
until the device appears.
until the device appears.
After the proper IPG is selected, click on the Connect button to start communication. The application will
then display the Main Screen.
then display the Main Screen.
The IPG is automatically interrogated by the Software Application so that the current device settings are
always available.
always available.
If an IPG error is encountered, a device error status will be displayed.
An automated adjustment of the IPG day clock takes place at the beginning of a session. The software sets
the time of day clock/calendar of the IPG equal to the computer time of day clock/calendar.
the time of day clock/calendar of the IPG equal to the computer time of day clock/calendar.