Understanding Test Results and Messages
The 
 Meter displays results between 10 and 600 mg/dL.
If the test result is below 10 mg/dL , “      ” will appear on the screen. Please 
repeat your test with by a new test strip.
If you still get a “      ” result, contact your healthcare professional.
If the test result is above 600 mg/dL, “      ” will appear on the screen. Please 
repeat your test again with a new test strip. 
If you still get a “      ” result, contact your healthcare professional.
GE006 Blood Glucose
About Quality Control Testing
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the 
 Control Solution GC550 when testing your GE006 Blood Glucose Monitoring 
System in the Control Solution Mode. If the test result is within the Control Solution Range printed on the 
strip vial label, the GE006 Blood Glucose Monitoring System passes the Quality Control Test and your 
GE006 Blood Glucose Monitoring System is working properly.
Rightest™
Control Solution Range:
Example of Control Solution Range printed on 
your test strip vial label.
Control Solution Range mg/dL
Level 2:              107-145 
Level 4:              315-426 
NOTE
- Level 2 of 
 is provided in your starter kit. If you want to 
purchase 
 or another 
 control solutions, 
Rightest™ Control Solution GC550
Level 4
Level 2
please contact Bionime Customer 
Service at 886 0800 241 688.