Bionime Corporation GM232 Manuale Utente

Pagina di 43
39
38
About Quality Control Testing
When should a Quality Control Test Be Performed ?
- To ensure that your GE006 Blood Glucose Meter and GE006 Blood Glucose Test Strip are   
working properly.
- To confirm that you are following the correct testing procedures.
- To prepare for your initial blood glucose test.
- To check the GE006 Blood Glucose Test Strip when you open a new vial of strips.  
- To  practice testing.
Required Items for Quality Control Tests
To perform a quality control test, prepare the items below :
- GE006 Blood Glucose Meter 
- GE006 Blood Glucose Test Strips
 
To check your GE006 Blood Glucose Meter after it has been dropped, damaged or exposed to liquids.
- If you suspect that your test results are inaccurate, or if your test results are not consistent with the 
way you feel.
  Control Solution GC550
Rightest™
CAUTION
- Each time you open a new bottle of control solution, write the expiration date on the label. 
The 
 is good for 3 months after opening the bottle, or 
until the expiration date printed on the label, whichever comes first.
Rightest™ Control Solution GC550
- Wipe the bottle cap with a clean tissue before tightly closing the bottle of control solution.
- Close the bottle of control solution tightly immediately after using.
- Check the expiration date before use. (Refer to the 
 
Package Insert).
- Keep control solution bottles out of reach of children.
Rightest™ Control Solution GC550
Example
Exp. Date 2014-11
About Quality Control Testing