TESTBOY TV 450 
Insulation resistance
 
31 
Insulation resistance measuring procedure 
 
* model TESTBOY TV 450 (TYPE B) 
‰
 
Select the 
INS
 function using the function selector switch. 
‰
 
Set the required test voltage
‰
 
Enable and set limit value (optional). 
‰
 
Disconnect tested installation from mains supply (and discharge insulation as 
required). 
‰
 
Connect test cable to the instrument and to the item to be tested (see figure 5.6).
‰
 
Press the TEST key to perform the measurement (double click for continuous 
measurement
 
and later press to stop the measurement). 
‰
 
After the measurement is finished wait until tested item is fully discharged. 
‰
 
Store the result by pressing the MEM key (optional)*. 
 
 
Figure 5.7: Example of insulation resistance measurement result 
Displayed results:  
R.........................Insulation resistance 
Um......................Test voltage – actual value.