Laser beta lasermike LS8000-3 Manuale Utente

Pagina di 221
LaserSpeed 8000-3 Instruction Handbook
 
Installing the System 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 17 of 221 
Revision A (Sep 2007) 
Mounting Distance 
When the material is within the depth of field, the gauge makes accurate 
measurements. Since the best signal comes from the center of the depth of 
field, the center is the source of the best measurements. This is because the 
two laser beams totally overlap at the center of the beam crossings and all of 
the laser power is used to form fringes. 
At the edges of the depth of field, however, some beam-crossing areas 
overlap and some do not. The areas that overlap form fringes and thus yield 
useful Doppler signals; the areas that do not overlap do not form fringes; 
rather, they produce only noise with no Doppler signals. Any measurements 
made outside the depth of field may not be reliable. 
Laser Safety 
When mounting of the gauge, ensure that the laser beams are blocked by 
machinery or beam blocks when material is not present in the measurement 
area. 
Environmental Conditions 
 
Ambient Temperature 
All LaserSpeed gauges have a built-in cooling system for applications in which 
ambient temperatures exceed 45ºC. 
Cooling fluids can be routed through the gauge using the 1/8-inch NPT fittings 
(1/4" NPT on LS8000E and LS8000X) on the back of the gauge. Typical flow 
rates of 1.5 litres/minute for water and 50 liters/minute for air are sufficient.  
There is minimal pressure drop through the gauge, so supply pressure can be 
very low. 
Ensure that the water temperature is above the condensation temperature or 
dew point, to prevent condensation from collecting on the front window of the 
gauge. If water condenses on the optical window, the laser beams could be 
blocked, resulting in reduced performance or loss of function. 
Coolant 
In applications with coolants on the material, an air jet may be required to 
remove the coolant from the measurement area. Since the speed of the 
coolant is generally different from that of the material, erroneous 
measurements could result. When incorporating an air jet, be sure to blow the 
coolant from the measurement volume at 90° from the material’s motion. If the