Installing the System 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 18 of 221 
Revision A (Sep 2007) 
coolant is blown off in the same direction the material is moving, the air 
velocity can add or subtract from the material’s velocity; this too may cause 
erroneous measurements. 
Air Wipe / Quick Change Window 
To prevent dust, dirt, and moisture from accumulating on the optical window 
the gauge can be fitted with an Air Wipe/Quick Change Window.  Clean, 
instrument grade air (see 
Specifications
 section for details) is blown at low 
pressure over the window to keep it clean.  The window can be removed for 
easy cleaning.  A clean, soft cloth should be used to clean the window.  If an 
abrasive cloth is used, the optical coating may be damaged, leading to poor 
measurements.  When replacing the window, put a small amount of vacuum 
grease on the O-Ring. 
 
 
 
 
 
 
 
 
 
 
For measurements and details about removing the Air Wipe/Quick Change 
Window, see 
Appendix A
 
Steam & Mist 
Steam or mist is sometimes present in some applications. The air-wipe is not 
designed to clear steam or mist between the gauge and material, but only to 
keep contaminants from depositing on the optics window.  Another means of 
removing steam or mist may need to be used. The steam or mist may be light 
enough to allow the laser beams to pass through without interference. If this is 
not true, a fan may be used to clear out the steam or mist between the gauge 
and housing. If a fan is not adequate or practical, an “air purge” can be 
installed to clear the beam path. Contact your Beta LaserMike representative