System Messages
12-4
Troubleshooting
6S2

&DEOH2II
The SpO
2
 cable is not connected to the 
device.
Attach the SpO
2
 cable to the HeartStart 
XLT.
632

)DLOXUH
A failure has occurred in the SPO
2
circuitry.
Remove the device from active use and 
call for service.
6S2

/LJKW,QWHUI
The level of ambient light is so high that 
the sensor cannot obtain an SpO
2
 reading 
or the sensor or cable is damaged.
z
Cover the sensor with an opaque 
material.
z
Check the sensor for damage; try 
another sensor.
6S2

1RQ3XOVDWLOH
The patient’s pulse is absent or too weak 
to be detected.
z
Check that the sensor is applied prop-
erly. 
z
Make sure the sensor site has a pulse.
z
Relocate the sensor to a site with 
improved circulation.
z
Try another sensor.
6S2

/RZ6LJQDO
SpO
2
 signal is too low to give an accurate 
reading.
z
Check the sensor is applied properly.
z
Try another sensor type.
6S2

1RLV\6LJQDO
Excessive patient movement, electrical 
interference, or optical interference is 
present.
z
Minimize patient movement or apply 
the sensor to a site with less move-
ment.
z
Secure the sensor cable loosely to the 
patient.
z
Reduce sources of electrical or optical 
interference.
6S2

6HQVRU)DLO
The SpO
2
 transducer is broken.
Try another sensor.
 
Table 12-1  System Messages  (Continued)
Message
Description
Corrective Action