System Messages
12-6
Troubleshooting
'DWD&DUG)XOO
z
The incident is more than 2 hours in 
duration, causing the Data Card to fill.
z
An empty Data Card was not inserted 
for the incident, causing the Data Card 
to fill sooner.
z
None. A new Data Card can not be 
inserted during an incident. 
z
Use one Data Card per incident/
patient to decrease the chance of the 
card filling.
'DWD&DUG,QWHUUXSWHG
The Data Card is removed during an inci-
dent.
z
None. The Data Card can not be re-
inserted during an incident.
z
Do not remove the Data Card during 
an incident.
'DWD&DUG1RW,Q6HUYLFH
The Data Card is inserted while the Heart-
Start XLT is on. 
None. A Data Card must be inserted prior 
to turning the HeartStart XLT on for the 
current patient. 
,QFRPSDWLEOH'DWD&DUG
A Data Card other than the M3510A is 
inserted.
Use only M3510A Data Cards.
1R'DWD&DUG3UHVHQW
A Data Card is not in the HeartStart XLT.
Turn the HeartStart XLT off and insert a 
Data Card prior to the first event for the 
patient. 
.H\,QDFWLYH
The key pressed is currently inactive (i.e. 
 
and 
 are inactive in 
AED Mode).
Use the appropriate mode for the key.
6WRS3DFHU
 is pressed while pacing pulses 
are being delivered. 
Stop pacing before changing the pacing 
mode.
Table 12-2  Momentary Messages (Continued)
Message
Possible Cause
Corrective Action
Sync
Pacer
Mode